Tech Note – Genomic DNA fragmentation using Covaris g-Tubes

Covaris g-Tubes can be used to fragment DNA and are recommended for use in creating Oxford Nanopore (ONT) libraries. The volumes and DNA concentration typically used for ONT libraries are outside the range of the example data provided by Covaris. Here, we tested using the volumes and concentrations appropriate for our requirements.

by Nick Crang and Orlando Contreras-López

To cite this study please see https://doi.org/10.1101/2024.05.29.596380

Covaris g-Tubes can be used to fragment DNA and are recommended for use in creating Oxford Nanopore (ONT) libraries. The volumes and DNA concentration typically used for ONT libraries are outside the range of the example data provided by Covaris. Here, we tested using the volumes and concentrations appropriate for our requirements.

Introduction

When making ONT libraries, it is sometimes beneficial to fragment the genomic DNA. Fragmenting the DNA will increase the molarity of the sample, along with making the length distribution more uniform, allowing for a more accurate estimation of sample molarity. The decrease in fragment length also increases the yield of the ONT library preparation, as longer fragments are more easily lost during clean up steps. This is compounded by the increased ease of sequencing of shorter fragments, which will result in a greater sequencing yield for the library. However, fragmentation does decrease the N50 of the library (defined as the largest length L such that 50% of all nucleotides are contained in contigs of size at least L1) and the length of the sequence reads is one of the great virtues of ONT sequencing.

There are a number of ways to fragment DNA for this, but one of the quickest is through use of g-Tubes (Covaris). Here, DNA is fragmented by passing the sample through a fixed aperture membrane using bench top centrifugation, resulting in DNA fragmentation with predictable fragment sizing. The relative centrifugal force (RCF) used to pass the sample through the aperture can be varied, resulting in greater or lesser mechanical shearing force upon the DNA and thus fragmenting it into shorter or longer fragments.

According to Covaris, three factors that can affect the fragmentation are: 

  1. Sample volume
  2. DNA concentration 
  3. RCF applied to the sample

It is worth noting that when fragmenting DNA to sizes <8 Kbp the processing time is reduced to 30 seconds from 1 minute. This is not discussed in the literature available from Covaris, and since this fragment size is much shorter than the 20 Kbp ideal we aim to achieve, the decision was made to maintain a 1-minute processing time for all samples. 

Rotary speed (RCF)
Targeted size
Mass of DNA6 Kbp8 Kbp10 Kbp20 Kbp
≤4 µg11 9045 1003 5421 735
8 µg11 9045 1003 5422 267
15 µg17 1427 2765 1003 310
30 µg17 1429 8387 2765 100
Processing time (seconds)30 sec60 sec60 sec60 sec

Table 1 A table provided by Covaris demonstrating the RCF required to fragment DNA to different lengths for different DNA concentrations. These values are for a sample volume of 150 µL.

NOTE: The original table was provided in RPM, this has been converted to RCF for ease of comparison using the formula RCF = 1.118 * 10^(-5) * r * RPM^2

We normally have a concentration of DNA around 1-2 µg provided by users for preparation of libraries for ONT. The volume we used for fragmentation is typically 50 µL rather than the 150 µL suggested by Covaris as this means we do not have to include a DNA concentration step to reduce our fragmented DNA to a usable sample volume for library input. We have previously observed this smaller volume and DNA concentration has a measurable effect on fragmentation, with an RCF of 1500 resulting in an average fragment size of between 14 051 bp when using 50 µL of 0.5 µg sample and 15 651 bp when using 50 µL of 1 µg sample. Both values are considerably smaller than the 20 Kbp suggested by Table 1. As the table provided by Covaris gave unified recommendations for samples containing  <4 µg DNA, we performed a fragmentation test using multiple concentrations within our expected sample range (0.5 and 4 µg).

In this study, we have two primary aims:

  1. Identify the RCF(s) most appropriate to achieve 20 Kb fragmentation of DNA within our operating range (0.5–4 µg in 50 µl)
  2. Identify the most appropriate method for size determination of large fragments (DNF-464 HS Large Fragment Kit for Fragment Analyzer or FP-1002 Genomic DNA 165 kb for Femto Pulse) 

We have additionally determined that an individual g-Tube can be reused multiple times, though membrane clogging typically becomes a problem after four sample preps.

It was decided to test three replicates each for the DNA quantities 0.5, 1, 2 and 4 µg at 1000, 1500 and 2000 RCF. All fragmentation tests were performed for 1 minute in a 50 µL volume. The degree of fragmentation was measured in both the Femto Pulse (Agilent), hereafter termed FEMTO, and the 5200 Fragment Analyzer System (Agilent), hereafter termed FA instruments. The kit used with the FEMTO should have greater resolving power for fragments above 20 Kb. The FEMTO has both a short and extended run settings which can be used to enhance the sample resolution, though this increases the run time from 40 minutes to 3 hours per sample.

Results

The sample preparation via the g-Tubes was conducted without incident, and fragmentation was clearly visible on the traces from both the FA and FEMTO (Figure 1). Prior to fragmentation, the trace was characterised by a broad peak with a very long tail up to the resolution limit of 165 000 bp. By contrast, after fragmentation a single clean peak was visible with a limited tail ending well below 50 000 bp. This pattern was consistent across all centrifugation speeds and quantities of DNA fragmented.

A problem was encountered when running samples on the FA, with numerous artefacts present during analysis, making it difficult to accurately size the fragments (Supplementary Figure 1). These artefacts appeared in multiple runs at different sizes and lanes, suggesting that they are poorly resolved DNA fragments rather than sample contamination. These artefacts were also seen when using the short run settings on the FEMTO, but were successfully resolved during an extended run. While there are only minor differences overall in the observed fragment sizes between the two instruments, the lack of artefacts from the extended run on the FEMTO gives us greater confidence in the accuracy of the results. The variation between replicates demonstrates the limitations of the g-Tube in terms of accurate fragmentation, with a size range of +/-4 Kb seen between replicates regardless of RCF and DNA quantity. Overall, when sizes could be resolved they were very similar between the FA and FEMTO with an R2 value calculated at ≥ 0.8. 

Figure 1 – Representative traces from extended run Femto Pulse. A) Shows unfragmented DNA, diluted to 0.5 µg / 50 µL. The tail of the peak extends well above 50 Kb and down to 1.3 Kb. The peak height is 42 Kb. B) Shows the DNA taken from the original 0.5 µg / 50 µL pool after fragmentation at 2000 RCF. A single clean peak is now visible with a peak height of 9 Kb and a tail reduced to significantly less than 50 Kb. C) Shows the DNA taken from the original 0.5 µg / 50 µL pool after fragmentation at 2000 RCF run on the FA. The high molecular weight artefact gives a much greater average fragment size despite a peak height of 12.5 Kbp.

We observed a general trend held true, that as the amount of DNA in the sample increased, the average size of the fragments increased for a given RCF (Figure 2). This effect was most apparent at 1000 and 1500 RCF, though it decreased as the RCF increased, with little difference between the four DNA concentrations at 2000 RCF. It was also observed, as expected, that increasing the RCF reduced the average fragment size. This effect was much more pronounced as the DNA concentration increased, with little difference seen between the three RCFS at 0.5 µg of DNA, where all nine samples were in the range 12-21 000 bp. By contrast at 4 µg the average size of each fragment at the three RCFs were separated by 10 000 bp, spanning the range 15-35 000 bp.

Fragmentation results measured with an extended run on the Femto Pulse # M5330AA (Agilent). Values are an average of three replicates with error bars of standard deviation. The three series indicate the centrifugation speed used during the DNA fragmentation measured in RCF.

Figure 2 – Fragmentation results measured with an extended run on the Femto Pulse # M5330AA (Agilent). Values are an average of three replicates with error bars of standard deviation. The three series indicate the centrifugation speed used during the DNA fragmentation measured in RCF.

Results from FEMTO
RCF
Amount of DNA (µg)1000 RCF1500 RCF2000 RCF
0.517 44516 75611 800
0.514 53613 21611 917
0.521 40015 80714 258
121 02218 46114 182
120 26616 11213 370
119 38814 79212 901
229 75221 56617 358
228 41118 36413 455
228 43020 36512 931
435 25225 22516 784
435 60624 08213 642
433 43423 31214 179

Table 2 – Fragmentation results measured with an extended run on the Femto Pulse # M5330AA (Agilent). Highlighted in blue text are RCF conditions that gave optimal fragmentation for that DNA concentration. No consistent difference was observed for fragmentation at 2000 RCF, or when the amount of DNA was 0.5 µg, so no values are highlighted.

A clear change was apparent in the traces after g-Tube fragmentation. The broad peak characteristic of the unfragmented DNA (Figure 2 A, Figure 3 A) was replaced with a single, far narrower peak (Figure 3 B-E). The breadth of the peak as well as the average fragment size increased along with the amount of DNA in the sample.

Figure 3 – Representative traces of the fragmentation results at RCF 1000 from 0.5 – 4 µg of input DNA. Samples measured with an extended run on the Femto Pulse # M5330AA (Agilent). A) Shows unfragmented DNA, diluted to 0.5 µg / 50 µL. B) Shows DNA diluted to 0.5 µg / 50 µL fragmented at RCF 1000. C) Shows DNA diluted to 1 µg / 50 µL fragmented at RCF 1000. D) Shows DNA diluted to 2 µg / 50 µL fragmented at RCF 1000. E) Shows DNA diluted to 4 µg / 50 µL fragmented at RCF 1000.

Discussion

On average the correlation between FA and FEMTO results is good (R2≥ 0.8), but individual measurements can differ as much as 5000bp between the systems. Based on this observation, if accuracy is needed we would recommend having replicates of the same sample in one run or running the same samples several times. Given the prevalence of unresolvable artefacts during FA and short FEMTO runs we would also recommend performing an extended run on the FEMTO to guarantee high quality usable results. If an approximation of the size is sufficient, then a short FEMTO run or a FA run are equally applicable.

Generally, smaller fragment sizes are given by FEMTO than FA, but this was not always the case. The most pronounced effect was that larger fragments tended to appear even larger when run on FA, which makes sense given the greater resolving power of the kit used with the FEMTO in this range.

It should be noted that sequencing performance of DNA fragmented this way has not been tested during this investigation. Anecdotal reports from other sequencing labs indicates that many more short fragments are left in the samples after using the g-Tube compared to the alternative fragmentation method Meagruptor 3 (Diagenode), suggesting a bead clean up step should be utilised if g-Tube fragmentation is chosen.

Overall based on our investigation it appears that g-Tubes can provide reasonably consistent DNA fragmentation and the fragment size can be varied in a predictable manner by varying the RCF used and the quantity of DNA. To achieve DNA fragments of approximately 20 Kb, for samples ≤ 1µg, 1000 RCF would be appropriate, whereas for 2 µg 1500 RCF would be best. Samples of 4 µg would likely need an RCF between 1500 and 2000 as the former resulted in fragments around 25 Kb and the latter around 15 Kb (Table 2).

Methods

Dilution

Commercially sourced human DNA #G1471 (Promega) was diluted using nuclease-free water to prepare four sample dilution pools containing 0.5, 1, 2 and 4 µg of DNA / 50 µL of volume. Each of these pools were 500 µL in total volume, and were used to provide replicate samples for each test condition. Pipette mixing of the pools was performed using wide bore pipette tips to prevent the high molecular weight DNA being damaged.

DNA fragmentation

A total volume of 50 µL was loaded into a single g-Tube #520079 (Covaris). Each g-Tube was used for a single RCF and reused four times, with samples being loaded in the order 0.5, 1, 2 and finally 4 µg of DNA. Fragmentation was achieved via one minute centrifugation at the specified RCF. Samples were run on a Centrifuge 5430 (Eppendorf).

Fragment Analyzer

Samples were diluted to within the working concentration of the assay with nuclease free water. Two µL of each diluted sample were run using the HS Large Fragment 50kb Kit #DNF-464-0500 (Agilent) according to the manufacturer’s instructions on the 5200 Fragment Analyzer System # M5310AA, 12 Capillaries, 33 cm. Data was analysed using PROSize v5.0 software (Agilent).

Femto Pulse

Samples were diluted to within the working concentration of the assay with nuclease-free water. Samples were run in both short and extended run formats on the Femto Pulse System, F12218, 12 Capillaries, 33 cm #SN F12218). Two µL of each diluted sample were run using the gDNA 165kb Analysis Kit #FP-1002-0275 (Agilent) according to the manufacturer’s instructions. Data was analysed using PROSize v5.0 software (Agilent).

References

  1. International Human Genome Sequencing Consortium. Initial sequencing and analysis of the human genome. Nature 409, 860–921 (2001). https://doi.org/10.1038/35057062

Acknowledgements

Our thanks go to Alfred Kedhammar for conducting an effective pilot study which informed our experimental design for this investigation.

Supplementary

Supplementary Figure 1 – Representative traces of the fragmentation results at RCF 1000 from 0.5 – 4 µg of input DNA. Fragmentation results measured on the 5200 Fragment Analyzer System # M5310AA (Agilent). A) Shows unfragmented DNA, diluted to 0.5 µg / 50 µL. B) Shows DNA diluted to 0.5 µg / 50 µL fragmented at RCF 1000. C) Shows DNA diluted to 1 µg / 50 µL fragmented at RCF 1000. D) Shows DNA diluted to 2 µg / 50 µL fragmented at RCF 1000. E) Shows DNA diluted to 4 µg / 50 µL fragmented at RCF 1000.

Last Updated: 30th May 2024